Multi-Hipot Tester
|
|
Chroma Model 19052/19053/19054 Multi-Hipot Tester |
Applications
|
Complete Dielectric Testing Solution
The Chroma Hipot Tester 19050 series
provide 3 models for choice. The 19052 for AC/DC/IR Hipot testing and
Insulation Resistance (IR) measurements, the 19053 which combines both AC and DC
Hipot tests and IR measurements with 8HV scan channel capability
into a single compact unit, and the 19054 which combines both AC and DC
Hipot tests and IR measurements with 4HV scan channel capability
into a single compact unit. Digital display and user friendly control
allows test parameters and limits to be set easily without the
high voltage activating.
The 19050
series electrical safety tester are advanced digital hipots with
load and line regulation to ensure the measurement integrity.
Multi-step capability allows users to perform multiple tests in
a sequence such as AC hipot followed by IR.
The 19052 AC/DC/IR Hipot Tester
performs AC/DC dielectric withstand (hipot)
tests. The test of AC voltage can be programmed in the range of
50V AC to 5kV AC with a resolution of 2V. Its maximum total output
current is 30mA. The test of DC voltage can be programmed in the
range from 50V DC to 6kV DC with a resolution of 2V. The maximum
total current is 10mA. The IR measurement range is from 1MΩ
to 50GΩ with test voltages from 50 to 1000V DC.
The 19053 AC/DC/IR/SC Hipot Tester
has all of the features of the 19052 plus 8HV Scan
Channels for multi-point testing.
The 19054 AC/DC/IR/SC Hipot Tester
has all of the features of the 19052 plus 4HV Scan
Channels for multi-point testing.
Quick Discharge
In DC hipot and IR tests the device under
test is discharged back through the HV transformer. This technique
results in a rapid and safe discharge.
Ground Continuity Check
All of the 19050 series testers have
a ground continuity check feature to determine the resistance,
that is between the ground blade of power cord and any exposed
metal on the product, is less than 1Ω.
Ground Fault Interrupters (GFI)
GFI is required by the National Electrical Code in wet locations. Such devices
automatically interrupt power when a ground current > 0.5mA exists
for more than a few milli-seconds to protect users.
Specifications | |||
Model | 19052 | 19053 | 19054 |
Mode | ACV/DCV/IR | ACV/DCV/IR/SCAN | |
Withstanding Voltage Test | |||
Output Voltage | AC : 0.05 ~ 5kV, DC : 0.05 ~ 6kV | ||
Load Regulation | 1% + 5V | ||
Voltage Resolution | 2V | ||
Voltage Accuracy | 1% + 5 counts | ||
Cutoff Current | AC : 30mA, DC : 10mA | ||
Current Resolution | AC : 1µA, DC : 0.1µA | ||
Current Accuracy | 1% + 5 counts (5% + 20 counts for real current) | ||
Output Frequency | 50Hz/60Hz | ||
Test Time | 0.3 ~ 999 sec., continue | ||
Ramp up Time | 0.3 ~ 99.9 sec., off | ||
Waveform | Sine wave | ||
Insulation Resistance Test | |||
Output Voltage | DC : 0.05 ~ 1kV | DC : 0.05 ~ 1kV | |
Voltage Resolution | 2V | 2V | |
Voltage Accuracy | 1.5% + 5 V | 1.5% + 5V | |
IR Range | 0.1MΩ~ 50GΩ | 0.1MΩ~ 10GΩ | |
Resistance Resolution | 0.1MΩ | 0.1MΩ | |
Resistance Accuracy |
≥ 500V
: 1MΩ~1GΩ:
±5% + 10 counts, 1GΩ~10GΩ: ±10% + 10 counts 10GΩ~15GΩ: ±15% + 10 counts, < 500V : 0.1MΩ~1GΩ: ±10% +10 counts |
||
Scanner Unit | -- -- | 8 ports, ±phase | 4 ports, ±phase |
Arc Detection | |||
Setting Mode | Programmable setting | ||
Detection Current | AC : 1mA ~ 15mA, DC : 1mA ~ 10mA | ||
Minimum Pulse Width | 10µs approx. | ||
Secure Protection Function | |||
Fast Output Cut-off | 0.4ms after NG happen | ||
Fast DC discharge | 0.2 sec. | ||
Ground Fault Interrupt (GFI) | 0.5mA ± 0.2mA AC (Low), 5mA ± 2mA AC (High), Close | ||
Panel Operation Lock | Present password | ||
Continuity Check | Current 0.1A, 1Ω ± 0.2Ω, Off | ||
GO/NG Judgment Window | |||
Indication, Alarm | GO: Short sound, Green LED; NG: Long sound, RED LED | ||
Data Hold | Least tests data memories | ||
Memory Storage | 99 steps or 99 groups for total 500 memory locations | ||
Remote Connector | |||
Real Panel 9 pin D-type connector | Input : Start, Stop, Interlock (at
11 pin terminal block only) Output : Under test, Pass, Fail |
||
General | |||
Operation Environment | Temperature : 0 ~ 40°C, Humidity : ≤ 80% RH | ||
Power Consumption | No load : < 100W. With rated load : 500W max. | ||
Power Requirements | 100V/120V/ 220V AC ±10%, 240V AC +5%~ -10%, 50/60Hz | ||
Weight | 14kg | 15kg | 15kg |
Dimension(W X H X D) | 320 x 105 x 400mm |
Model | Description |
---|---|
19052 | AC/DC/IR Hipot Tester |
19053 | AC/DC/IR Hipot Tester 8ch Scan |
19054 | AC/DC/IR Hipot Tester 4ch Scan |
A190501 | Manual Operation TR. Scan Box |
A190512 | Auto Control TR. Scan Box |
A190513 | Auto Control Dual TR. Scan Box |
A190508 | GPIB Interface |
A190510 | Printer Interface |