Multi-function Gain-Phase Analyzer
Model GP102
|

|
Typical applications
Testing of closed loop feedback systems
Synchro, resolver, RVDT and LVDT test
Inductor / transformer testing
Incoming inspection of components
Complex impedance measurement
Core loss measurements
Ultrasonic power measurements
Power line carrier signal measurements
Multi-function testing in University labs
|
Description
The Model GP102 is a multi-function instrument which
is ideal for users that must make a large variety of sophisticated measurements. A single
package, contains a Gain Phase Analyzer, Phase Angle Voltmeter with a null detector,
Function Generator, Pulse Generator, Spectrum Analyzer, Digital Sampling
Oscilloscope, Wattmeter, Phase Meter and Complex Impedance Bridge.
These internal instruments cover the frequency range from dc (10mHz for the ac
instruments) to more than 2MHz.
Each of these internal instruments may be used independently or in an
internally correlated fashion. By selecting the internal Function Generator and the two
Voltmeters, the user can obtain dual gain-phase data for two outputs of any network he is
testing. This is particularly useful for making Bode plots to check the stability of
networks having feedback. These plots can be displayed in several formats using the
oscilloscope capability of the Model GP102. Because of the floating inputs, the
measurements can be made across network components with neither side connected to ground.
In its Phase Angle Voltmeter (PAV) mode, the GP102 presents, on a
single display, the total rms, the in-phase and quadrature components, the phase angle and
the fundamental magnitude. An additional display provides the ratios of the various
parameters. An internal null meter provides the user with the ability to make
precision ratio adjustments. The specifications in this mode are superior to the
specifications of all competing PAVs regardless of their cost. In addition, unlike
traditional PAVs, the GP102 has the advantage of a built-in signal generator. All of these
features make the GP102 ideal for synchro, resolver, RVDT and LVDT testing.
The Spectrum Analyzer feature can be used to observe the harmonic content
of complex, periodic input signals. This proves to be especially useful for examining
current waveforms to ensure that they do not exceed the requirements placed on harmonics.
As a true rms Wattmeter, the Model GP102 is ideally suited to measure core
loss, input power to ultrasonic transducers, power consumed by switching power supplies,
and many other devices operating at higher frequencies. It also performs exceptionally
well with measurements at 50Hz or 60Hz.
The Complex Impedance Bridge function permits the user to measure
Resistance, Inductance and Capacitance at frequencies from dc (Resistance only) to 2MHz.
Secondary functions such as Q (quality factor), D (dissipation factor), and ESR
(equivalent series resistance) can also be measured over the same frequency range.
Data are presented on an orange, vacuum luminescent display either in
tabular form or in a swept graphical form, e.g., an amplitude vs. frequency plot for a
linear network. The tabular form comprises a list of the frequencies with their
corresponding amplitudes. A zoom function permits the user to enlarge items of interest on
the display.
Typical Displays
GP102 Specifications
CH1 & CH2 Voltmeters
| Channels / display |
2 isolated / 5 digits |
| Measurement |
Rms, ac, dc, peak, cf, surge |
| Frequency |
Dc to 2MHz |
| Coupling |
Ac or ac+dc |
| Max input |
±500V peak
±500V peak from earth |
| Input ranges |
500V 300V, 100V, 30V, 1 0V, 3V, 1V, 300mV, 100mV, 30mV,10 mV |
| Ranging control |
Full auto, up only, or manual |
| Input impedance |
1 M // 30pF (excluding. Leads) |
Accuracy (ac)
All + 0.4mV |
0.1 % range < 1 kHz
0.3% range < 10kHz
1 % range < 50kHz + 0.004% rdg per kHz to 2MHz |
| Accuracy (dc) |
0.2% range + 2mV |
| CMRR |
>55dB @ 10V 1 MHz
>110dB @ 100V 1 kHz |
| Time constant |
0.2s, 1.5s or 12s |
Phase Meter
| Frequency |
10mHz to 2MHz* |
| Measurement |
DFT analysis |
| Phase accuracy |
0.02° < 100Hz 0.05° < 1 kHz
0.25° < 10kHz + 0.005° per kHz |
* 20mHz to 1 MHz if not using generator
Gain phase analyzer
| Frequency range |
10mHz to 2MHz* |
| Modes |
Gain, angle, real, imaginary |
| Accuracy |
±0.02dB |
* 20mHz to 1 MHz if not using generator
LCR meter
| Frequency range |
10mHz to 2MHz* |
| Measurement |
DFT analysis |
| Functions |
L, C, R (ac), R (dc), Z, Q, tanð, Ø, series or parallel |
| Ranges |
10pF to 1F
100nH to 1000H
100mΩ to 10MΩ |
| Accuracy |
0.3% < 1 kHz, 0.75% < 10kHz 2.5% < 50kHz, 12.5% < 1 MHz |
* 20mHz to 1 MHz if not using generator
Transformer analyzer
| Inductance L |
100nH to 1000H Series / Parallel |
| Quality factor Q |
0.001 to 2000 |
| Leakage inductance |
100nH to 1000H |
| Inductance match |
1:5000 to 5000:1 |
| Turns ratio (voltage) |
1:70000 to 70000:1 |
| Turns ratio (Inductance) |
1:2000 to 2000:1 |
| Winding polarity / phasing |
Pass/fail or ±indication |
| Dc R & winding continuity |
20mΩ to 100kΩ |
| Inter winding phase angle |
0° to 360° or ±180° |
| Impedance test |
10mΩ to 100MΩ |
| Insertion / return loss |
-100dB to 100dB |
| Frequency response |
10Hz to 2MHz |
| Longitudinal balance |
0dB to 100dB |
| Inter winding capacitance |
10pF to 1F |
| Inter winding capacitance match |
1:5000 to 5000:1 |
| Distortion THD |
0.1 % to 100% |
Signal generator
| Waveforms |
sine, triangle, square, sawtooth, dc only |
| Frequency |
10mHz to 2.4MHz (sine); 10mHz to 1MHz (other) |
| Accuracy |
Frequency ±0.05%; Amplitude ±5% (to 100kHz) |
| Output impedance |
50Ω |
| Output voltage |
10mV to 10V peak |
| Offset |
0V to 10V |
Pulse generator
| Frequency |
10mHz to 2.4MHz |
| Pulse width |
200ns to 10s (50ns resolution) |
| Rise and fall time |
selectable 50ns (5V) to 1s |
Phase Angle Voltmeter PAV (Vector Voltmeter)
| Measurement |
Total rms, in-phase and quadrature components, phase angle, fundamental
magnitude, and ratio with null meter capability. |
| Range |
10mHz to 1MHz |
| Accuracy |
See voltmeter/phase meter specifications |
Low frequency DSO / waveform display
| Channels |
2 |
| Time-base |
20µs to 5s per division |
| Roll mode |
Time-base >1s/div |
| Pretrigger |
None, 25%, 50%, 75% |
| Sample rate |
800k sample/sec |
Power meter
| Measurements |
W, VA, power factor, V, A total, fundamental, integrated |
| Current and Voltage accuracy |
±0.1 % reading |
Power accuracy (power factor >0.7) |
±(0.3% rng + 0.3% rdg), f < 1kHz;
±(2% mg +0.5% rdg), f < 50kHz;
±(2% rng + 5% rdg), f < 1MHz |
Selective level meter
| Frequency range |
10Hz to 2MHz |
| Scan |
single, dual, or sweep |
| Selectivity (-3dB) |
3Hz, 24Hz, or 100Hz |
General
| Display |
160 x 80 dot graphic electroluminescent |
| Size |
Approximately 30 x 15 x 25 cm (Upright unit) |
| Temperature range |
Operating: 0° to 40°C
Within specification: 23° ±5°C |
| Weight |
Approximately 5kg |
| Power supply |
230 V rms ±10%, 50Hz, 30VA max. |
| Warranty |
1 year |
Inizio Pagina
|
|