M-509 Capacitance Decade
Capacitance decade is aimed for calibration of capacitance ranges of
universal multimeters and for calibration of service RCL meters.
However its usage is much more general. It can be used in for repair
centres, for developing purpose, etc. Decade is housed in metal
instrument case.
Technical data
| Capacitance range |
100 pF - 12,221 µF |
| Maximal voltage |
50 Vef |
| Connection |
Two-wire, ground terminal |
| Control |
Rotary switches |
| Dimensions |
W 325 mm, H 111 mm, D 260 mm |
Accuracy
| Decade |
Accuracy (%) |
Dissipation factor (-) |
Temperature coefficient (%/°C) |
| 100 pF - 1 nF |
2 |
< 0.001 |
< 0.01 |
| 1 nF - 10 nF |
1 |
< 0.001 |
< 0.01 |
| 10 nF - 100 nF |
0.5 |
< 0.005 |
< 0.02 |
| 100 nF - 1 µF |
0.5 |
< 0.005 |
< 0.02 |
| 1 µF - 11 µF |
0.5 |
< 0.005 |
< 0.02 |
|