DP6
Thermocouple Calibrator / Simulator
|
 |
DP6 Data Sheet Click to view in PDF format (106KB)
|
Linearised for 10 Thermocouple Types
Measures and Sources
Calibrated to ITS90 and IPTS68
Displays in °C, °F, K and mV
1µV Resolution
Reference Junction Compensation
Rechargeable Battery
Memory with 130 Stores
Digital Calibration
|
The DP6 portable thermocouple calibrator is a compact and easy to use
instrument that is equally at home in the laboratory or service environment.
The DP6 will both measure and simulate the outputs from the 10 most popular
thermocouple types, and may be scaled in °C, °F, K or mV.
Automatic reference junction compensation is provided with the option for
manual value entry.
For full portability the calibrator is powered by a rechargeable
battery giving more than 15 hours continuous use.
The DP6 is delivered in an ABS transport case which also houses the
charger and a copper transition adapter for easy connection to miniature
thermocouple plugs. A soft carrying bag with neck strap is available as an option.
DP6 Technical Specifications
| |
Thermocouple Type
|
Range
|
Accuracy
|
|
|
Thermocouple Type
|
Range
|
Accuracy
|
|
Code
|
Material
|
Degrees °C
|
Measure & Source
|
|
Code
|
Material
|
Degrees °C
|
Measure & Source
|
|
B
|
PtRh30-PtRh6
|
+500 to +1820
|
± 0.5 °C
|
|
N
|
NiCrSi-NiSi
|
+1100 to +1300
|
± 0.4 °C
|
|
B
|
PtRh300-PtRh6
|
+200 to +500
|
± 1.5 °C
|
|
N
|
NiCrSi-NiSi
|
+400 to +1100
|
± 0.3 °C
|
|
B
|
PtRh300-PtRh6
|
+60 to +200
|
± 6.0 °C
|
|
N
|
NiCrSi-NiSi
|
+150 to +400
|
± 0.15 °C
|
|
E
|
NiCr-CuNi
|
-200 to +1000
|
± 0.2 °C
|
|
N
|
NiCrSi-NiSi
|
0 to +150
|
± 0.1 °C
|
|
E
|
NiCr-CuNi
|
-250 to -200
|
± 0.6 °C
|
|
R
|
PtRh13-Pt
|
+1200 to +1760
|
± 0.8 °C
|
|
E
|
NiCr-CuNi
|
-270 to -250
|
± 6.0 °C
|
|
R
|
PtRh13-Pt
|
+100 to +1200
|
± 0.4 °C
|
|
J
|
Fe-CuNi
|
+800 to +1200
|
± 0.3 °C
|
|
R
|
PtRh13-Pt
|
0 to +100
|
± 0.5 °C
|
|
J
|
Fe-CuNi
|
+200 to +800
|
± 0.2 °C
|
|
R
|
PtRh13-Pt
|
-50 to 0
|
± 0.8 °C
|
|
J
|
Fe-CuNi
|
0 to +200
|
± 0.1 °C
|
|
S
|
PtRh10-Pt
|
+1400 to +1760
|
± 0.95 °C
|
|
J
|
Fe-CuNi
|
-210 to 0
|
± 0.3 °C
|
|
S
|
PtRh10-Pt
|
+1200 to +1400
|
± 0.5 °C
|
|
K
|
NiCr-NiAl
|
+1000 to +1370
|
± 0.4 °C
|
|
S
|
PtRh10-Pt
|
+50 to +1200
|
± 0.4 °C
|
|
K
|
NiCr-NiAl
|
+100 to +1000
|
± 0.3 °C
|
|
S
|
PtRh10-Pt
|
-50 to +50
|
± 0.6 °C
|
|
K
|
NiCr-NiAl
|
-50 to +100
|
± 0.1 °C
|
|
T
|
Cu-CuNi
|
-100 to +400
|
± 0.2 °C
|
|
K
|
NiCr-NiAl
|
-150 to -50
|
± 0.2 °C
|
|
T
|
Cu-CuNi
|
-230 to -100
|
± 0.5 °C
|
|
K
|
NiCr-NiAl
|
-225 to -150
|
± 0.5 °C
|
|
T
|
Cu-CuNi
|
-250 to -230
|
± 1.0 °C
|
|
K
|
NiCr-NiAl
|
-270 to -225
|
± 3.0 °C
|
|
T
|
Cu-CuNi
|
-270 to -250
|
± 2.5 °C
|
|
L
|
Fe-CuNi
|
+300 to +900
|
± 0.2 °C
|
|
U
|
Cu-CuNi
|
+300 to +400
|
± 0.2 °C
|
|
L
|
Fe-CuNi
|
-100 to +300
|
± 0.1 °C
|
|
U
|
Cu-CuNi
|
0 to +300
|
± 0.1 °C
|
|
L
|
Fe-CuNi
|
-200 to -100
|
± 0.15 °C
|
|
U
|
Cu-CuNi
|
-150 to 0
|
± 0.15 °C
|
| |
|
|
|
|
U
|
Cu-CuNi
|
-200 to -150
|
± 0.2 °C
|
Resolution on all types of thermocouple: 0.1°C, 0.1°K, 0.1°F.
Limits of error apply for 1 year at 20 degrees centigrade ± 1°C
|
RANGE
|
MAX. DISPLAY
|
UNCERTAINTY
|
RESOLUTION
|
|
10 mV
|
± 15.000 mV
|
± 0.02% of reading ± 0.015% FS
|
1 µV
|
|
100 mV
|
± 150.00 mV
|
± 0.01% of reading ± 0.015% FS
|
10 µV
|
|
1 V
|
± 1.5000 V
|
± 0.01% of reading ± 0.015% FS
|
100 µV
|
Temperature coefficient: typically 17 ppm/°C +0.2 mV /°C
Display
4 ½ digit high contrast LCD 10.2 mm. Display
range 19999 digits, automatic decimal point, polarity and units. Two
line alphanumeric LCD for programming and display of configuration.
Terminals
Two 4mm low thermal emf copper terminals.
Working Temperature
0…40°C.
Storage Temperature
-20 … +50°C.
Battery
6 Volt 1.2 Ah sealed lead acid, replaceable.
|
Operating Time
15 Hours typical continuous use.
Charger Type
External charger operating from mains supply.
Case
Shockproof thermoplastic with polycarbonate sealed
membrane keyboard.
Soft carrying case available as an accessory.
REFERENCE JUNCTION
Referenced to 0 Deg. C and with three operating modes. Automatic with
internal sensor, Off (=to 0°C), and Manual entry via keyboard.
|
Accuracy
Better than ± 0.1°C at +20°C.
Deviation
0.01°C / °C over the range 0 to +50°C.
Manual Input Range
The reference junction
reference value may also be set via the keyboard
over the range 0 to +100°C.
Size
150 x 130 x 60 mm (W D H)
350 x 260 x 65 mm in transport case.
Weight
1.4 kg .
|
Inizio Pagina
|