LCR Meter mod. 11022 / 11025
LCR Meter 11022
Chroma model 11022
LCR Meter








LCR 11022-11025 Data Sheet
Click to view in PDF format  (170KB)

   

  • Transformer test parameters (11025), Turns Ratio, DCR, Mutual Inductance
  • 50Hz, 60Hz, 100Hz, 120Hz, 1kHz, 10kHz, 20kHz, 40kHz, 50kHz, 100kHz test frequencies
  • 0.1% basic accuracy
  • 21ms measurement time in all test frequency range (≥100Hz)
  • HP 4263B LCR Meter compatible IEEE-488 commands
  • 4 different output resistance modes selectable for non-linear inductor and capacitor measuring. Measured results are compatible with Chroma (Zentech) and other well-known LCR meters
  • High resolution in low impedance(0.01mΩ) and high accuracy 0.3% till 100mΩ range are the right tool for low inductance, large capacitance, and low impedance component measuring
  • Adjustable DC Bias Current up to 200mA (Constant 25Ω), are the right tool for inductance inspection of telecom transformers and small power chokes under DC Bias Current (11025)
  • 1320 Bias Current Source directly control capability
  • 240 x 64 dot matrix LCD display, easy operation
  • 0.01mΩ ~ 99.99MΩ wide measurement range with 4 1/2 digits resolution
  • BIAS comparator function
  • Comparator function and 8/99 Bin-sorting function
  • Alarm for PASS/ FAIL judge result
  • 50 internal instrument setups memory
  • Handler interface trigger edge (rising/falling) programmable
  • Test signal level monitor function
  • Standard GPIB (IEEE-488) and handler interface
  • Open/ short zeroing, load correction
 


The Chroma 11022 and 11025 LCR Meters are the measurement instruments for passive components. They are applicable to the automatic manufacturers for passive components in material inspection. With the features of 21ms high-speed measurement and 0.1% accuracy, 11022 LCR Meter fulfills the requirements for fast production. Its functions of 8-level counting, 8/99 Bin-sorting, pass/fail judgment, and 50 sets of internal save and recall settings totally meet the production line requirements for easy operation.

The four impedance output modes can measure the results with the LCR Meters of other brands to get a common measurement standard. Chroma 11025 LCR Meter is compatible with HP 4263B LCR Meter IEEE-488.2 control interface and has three impedance output modes for selection. The measurement results can also be compared with other brand of LCR Meters. Chroma11022/11025 is the ideal selection for passive components quality assurance and automatic production.


 

Model      Description
11022  LCR Meter
11025  LCR Meter
A110104  SMD Test Cable #17
A110211  Component Test Fixture
A110212  Component Remote Test Fixture
A110232  4 BNC Test Cable with Clip#18
A110234  High Frequency Test Cable
A110236  Rack Mountain Kit
A110239  4 Terminals SMD Electrical Capacitor Test Box (Patent)
A110242  Battery ESR Test Kit
A110244  High Capacitance Capacitor Test Fixture